Implementation of a QC procedure based on the absolute dosimetry of 192Ir-HDR sources by micro-ioniziaton chamber
'
Implementation of a QC procedure based on the absolute dosimetry of 192Ir-HDR sources by micro-ioniziaton chamber
-
DOI: https://doi.org/10.22533/at.ed.317522503024
-
Palavras-chave: '
-
Keywords: '
-
Abstract: The purpose of this report is to establish a method for the direct measurement of the absorbed dose delivered by 192Ir high-dose-rate (HDR) brachytherapy (BT) sources using a PinPoint ionization chamber. After a literature review, a quality control procedure for the absorbed dose-in water has been developed and included in the quality assurance (QA) of the brachyhterapy unit GammaPlus iX (Varian Medical Systems, Palo, CA).
- Miguel Martinez-Albaladejo